PER CONDITIONING PROTOCOL


The purpose of this document is to standardize the training
protocols for Proboscis Extension Response conditioning of honey bees.
Anyone performing PER studies should be aware of how variations in the
training procedure can affect the results. Indeed it is impossible to completely
standardize the protocol across all experiments and researchers, because
many times experimental manipulations require that the protocol be changed.
Nevertheless, there are aspects of the procedure that need to be used in
the proper way in order to minimize the possibility of artifact. At least
then any variation in the procedure would change as few variables as is
necessary across experimenters. The document that follows is meant as a
beginner's guide to the PER protocols. It was last updated on 3-May-1998.
Click on any of the following links to get to that part of
this page, then click on the 'back' button of your browser to return to
this location.
PIO12 PROGRAM
AND RELAY BOARD
DELIVERY OF THE CS
DELIVERY OF THE US
RECORDING RESPONSES
EXAMPLES
OFRESPONSES TO
CONDITIONED ODORS
CONDITIONING
AND TEST TRIALS
MISCELLANEOUS STUFF
EQUIPMENT YOU'LL
NEED
CONSTANCY
-
Whatever procedure you settle on, run it by others in the
laboratory for comments
-
vary this
procedure within your experiment
-
if you do, the results may not be statistically comparable!!!
ODOR SETUP
-
Make sure that odor does not get on your hands during the
procedure
-
wear gloves if necessary
-
thoroughly wash your hands afterwards
-
We us 1cc glass tuberculin syringes for odor cartridges and
automate odor delivery via the PIO12 program (see below)
-
we no longer use manual or automated delivery from 20 cc
plastic syringes
-
Make sure the 1cc glass syringe is clearly labeled in regard
to the odor for which it has been used
-
an odor is defined as the volatile pattern given off by an
odorant (pure) or a mixture
-
label in writing and by color: 1-hexanol (green); geraniol
(yellow); 2-octanone (blue); hexanal (white); all mixtures (red); everything
else (orange)
-
Rinse the inside of the 1 cc glass tuberculin syringe with
alcohol and wash off the black rubber restrictor that goes into the end
of each syringe
-
Set up the syringe on the ledge of the exhaust hood
-
never open jars of odor anywhere but in front of this hood
in room 6
-
Place a small strip of filter paper into the tube such that
the end sticks out about 1-2 cm
-
Place 3 ml of pure odorant or 5 ml of a dilution in solvent
onto the filter paper and then discard the tip of the Pipetteman before
setting up the next odor
-
do not let the odor touch the end of the tube!
-
the Pipetteman tip can be reused for the SAME odorant, but
never stick it into two different odors
-
Turn the syringe upward so that the filter paper slides rapidly
into the tube
-
Thoroughly wash the tube with alcohol after setup to prevent
any odor from tip from reaching the bee
-
ball up the end of a Kimwipe soaked in alcohol and make sure
to thorougly wash the black plug in addition to the outside of the glass
part of the syringe
-
GENERAL RULES:
-
IF IN DOUBT about prior usage of a syringe, get a new syringe
and discard the old one
-
mix tubes or plugs; that is, the same plug should always go into the same
tube
-
once a syringe has been used for a particular odorant or
mixture it can only be used for that odor REGARDLESS of how much it has
been cleaned
-
Setup up all of the different odors you’ll need for acquisition,
and setup a NEW set of syringes for TEST trials
COLLECTION AND SETUP OF DRONES,
WORKERS AND QUEENS
-
WORKERS should be collected from the entrance or from frames
by placing a scintillation vial over them until they fly into it
-
make sure the lid has a hole in it!
-
only one bee/vial
-
place the vials into an ICE/WATER SLURRY until the bee inside
stops moving
-
remove immediately to avoid overexposure to cold and place
the bee into a harness
-
DRONES should be collected in vials late in the day as they
return from mating flights
-
put an excluder in front of the entrance and let returning
drones collect on it
-
place them into an excluder cage
-
put the cage into a cell builder overnight
-
make sure there is space between the cage and the frames
so that nurse bees can enter the cage to feed to drones
-
early the next day place the cage into a small chamber and
remove the plug
-
place the drones directly into harnesses
-
QUEENS should be removed from cages and placed directly into
restraining harnesses
-
Make sure the tape on the harness is tight
-
too much space and the bee will escape or frequently not
be able to extend its proboscis properly
-
There are two different times to do the setup
-
use one for each experiment!!
-
don’t switch between them within an experiment
-
you may need to use one or the other depending on whether
you’re using flightroom or outside bees, queens/drone/workers, etc.
-
if you’re comparing different groups of bees THEY NEED TO
BE TREATED IN THE SAME WAY!!!
-
METHOD 1: SETUP ON THE DAY OF TESTING:
-
setup bees in harnesses the first thing in the morning
-
between 30 min and 1 hr, no more, after setup feed all bees
0.4 ml of the sucrose to be used for the US on that day
-
allow at least 1 hr (preferably 2 hrs) beyond this point
for recovery before training
-
METHOD 2: SETUP ON THE DAY BEFORE TESTING
-
make sure the setup procedure is COMPLETE at least by 24
hr before testing is to begin
-
between 30 min and 1 hr, no more, after setup feed all bees
0.4 ml of the sucrose to be used for the US during training
-
feed on several different passes through a set of bees until
the bees are satiated
-
that is, until the proboscis is no longer extended by antennal
contact with sucrose
-
place bees on a countertop overnight
-
do not feed them the next morning
-
24 hr after feeding begin the training protocol
-
note that we generally find poorer conditioning performance
with this protocol, which probably stems from lower overall motivational
states of the bees
-
CAUTION: preexposure to the US can influence subsequent
conditioning!
-
therefore feed in an area well removed from the training
locale
PIO12 PROGRAM
AND RELAY BOARD
-
The PIO12 program is used to time the delivery of odors and
signal delivery of the US
-
The program will prompt you for a <filename>.stm file
when it begins
-
this file lists the different delivery protocols (up to 9)
for each experiment
-
make sure you are aware of the contents of the specific file
you are using!
-
At the beginning of each day, and maybe even at intervals
throughout each day, check the electrical and air connections
-
you can check air delivery by placing a moist fingertip in
front of the top outlet of the white valve on the relay board
-
if you don’t feel a strong air stream, then something’s wrong
-
check the exhaust system in the same way to make sure the
tubing has not come apart
-
also, if the electrical connection is broken for some reason,
the computer will work but the board and/or the valve will not
-
make sure the LED lights work on a trial run of the system
before you begin each day
-
also, if it’s not working, check to make sure that the switches
underneath the table in front of the hood are on and that the aquarium
pumps are on
DELIVERY OF THE CS
-
Adjust the odor syringe in the clay holder so that it points
directly at the bee’s antennae
-
use a small plexiglass stand on top of which some modeling
clay is placed
-
the glass odor syringe can be placed in the clay and adjusted
to point at the bee's head
-
Securely place the fitting connecting the syringe to the
valve over the ground glass end of the syringe
-
Position the end of the syringe 1-2 cm from the bee
-
no farther than 2 cm and not closer than 1!!
-
pick a distance within this range and be constant about it
-
If the bee begins to respond immediately to the syringe,
try to smell odor on the tip
-
if you can smell some when the other end is closed by the
fitting, wash it off with a Kimwipe soaked in 95% ethanol
-
if you continue to have trouble with the syringe throw it
out and get a new one
-
The air flow rate is on the order of 400 ml/min coming in
from the aquarium pumps
-
it is good to check this flow rate periodically
-
regularly (daily) make sure the valve is switching properly
by placing the tube coming out from the valve into a vial of water or hooking
it up to a flowmeter
-
400 ml/min means that about 7 ml/sec are pushed through the
odor cartridge; so over 4 sec 28 ml will be pushed through a 1 ml cartridge
DELIVERY OF THE US
-
Load the Gilmont syringe with the sucrose solution so that
there are no bubbles in the glass portion and with the micrometer wound
back as far as possible
-
Wind the micrometer forward a turn or two to make sure that
the fluid in the needle is actually sucrose and not water from a wash
-
Turn the micrometer to deliver a multiple of 0.2 ml, which
is the smallest increment on the syringe
-
we have fed a number of ways in the past, but it should from
now on be standardized
-
different protocols will require different amounts of feeding
-
generally we should be working in the 0.2, 0.4, 0.6 and 0.8
ml range
-
use “ad
libitum” feeding, which is a large droplet that the bee feeds from for
several seconds and cannot entirely consume
-
Hold the syringe approximately 2 cm from the bee as odor
is presented
-
get it too close in anticipation of
the feeding signal!!!!!
-
use the edge of the training arena to balance your hand so
that it does not shake
-
When the feeding signal is delivered by the computer LIGHTLY
touch the antennae until the proboscis extends, then feed the bee
-
this should require only a very light touch to the antennae
-
watch out for sucrose build up on the antenna
-
GET THE
SUCROSE TOO CLOSE BEFORE THE FEEDING SIGNAL OCCURS, and make sure you’re
there with it as soon after the beep as is possible
-
If you are using up to 0.8 µl the bee should easily
consume the entire droplet within the time period alloted for the US
-
Watch for buildup of sucrose on the antennae and modify your
procedure if you see it!
RECORDING RESPONSES
-
A response (‘+’) is defined as anytime that the bee extends
its proboscis beyond the line made by connecting the tips of the opened
mandibles
-
stay away from trying to score partial responses - if you
need to make finer distinctions use the video technique outlined below
-
only score a ‘+’ when the behavior begins with or after the
onset of odor
-
if the proboscis i never extended, or is extended after the
offset of odorant, then it is not a response (i.e., a ‘-’)
-
Scoring responses as either a + or a - is satisfactory for
many kninds of analyses
-
treatment groups pf 20-40 bees can be compared in regard
to the percentage of bees that responded in the test situation
-
However, much more information can be obtained by videotape
analyses of the response to an odor
-
I have several examples now in which the percent response
levels in two treatment groups are identical, but the groups differ in
other aspects of the response topologies (e.g. duration and latency)
-
For video analyses make sure that the recording time extends
for about 20 sec after onset of odorant, because a response that begins
during the odor presentatation can easily extend beyond the offset of odor
-
always use extinction (odor only) trials for video recording
-
that is, don’t record acquisition trials, because you can
only score latency in those kinds of trials (the response is contaminated
by the presence of food)
-
You can score many response measures from slow frame advance
mode on the videotape
-
duration = time that the proboscis broke the mandible line
(above) subtracted from the time that the line is again broken during retraction
-
multiple extensions/retractions can occur during the 20 sec
recording and they must be calculated individually, even though you may
ultimately decide to add all of them together to obtain the total time
the proboscis was extended during the trial
-
latency = time that the first extension occurs minus the
time the light (odorant) comes on
-
glossal extensions can be counted and timed relative to the
onset of odorant, the first extension or their own timing parameters
EXAMPLES
OF VIDEOTAPED RESPONSES TO CONDITIONED ODORS
-
The following links go to video sequences from extinction
(odor only) trials
-
note that in all three examples the bees respond with proboscis
extension to odor, so theyall would have received the same visual score
(a '+')
-
nevertheless, they all differ in terms of latency, duration,
and the number of extensions/retractions during a single 4 sec odor pulse
-
in each frame, the camera is focused directly at the front
'face' of the bee
-
the mandibles and proboscis are oriented toward the top of
the frame
-
the two antennae are also free to move
-
Each video frame contains the frame code in the lower right
portion of the frame
-
the number to the left of the colon indicates the number
of seconds into the tape at which the frame occurs
-
the number to the right of the colon indicates the number
of frames into that second
-
given that there are 30 frames/sec, you can convert to absolute
time by multiplying the latter number by 0.033 and adding it to the seconds
-
The red light at the bottom of the frame indicates that the
odor is being presented
-
bees cannot see red, and the light is placed behind and below
the bee
-
when this light comes on (the first frame in each sequence)
the time reference for the trial is set to 0
-
Also, note the antennal movements
-
this page focuses on proboscis extension as a measure of
the conditioned response
-
but the antennae also show characteristic odor-oriented movements
that can be conditioned
-
SHORT LATENCY/LONG DURATION
-
LONG LATENCY/SHORT DURATION
-
SHORT LATENCY/SHORT DURATION
CONDITIONING
AND TEST TRIALS
-
Educate yourself about Interstimulus Interval (ISI) and Intertrial
Interval (ITI) by reading Rescorla (Ann. Rev. Neurosci., 1988)
-
Describe conditioning and test trials
-
use different (new/charged) syringes for testing
-
the ones used for acquisition may be differentially depleted
if you've used them for different numbers of trials
MISCELLANEOUS STUFF
-
Remember: bees like it hot! We keep our training rooms about
82-85oF (29-30oC)
-
Contact Brian H. Smith (smith.210@osu.edu) for questions/problems
EQUIPMENT YOU'LL
NEED
-
I is VERY important to accuratly and precisely regulate the
delivery of the CS and the US
-
You will need a means to automate the odor delivery system
-
we use our own hardward that is driven by a parallel port
on a computer system
-
The parallel port controls a series of valves that regulate
the flow of air through the odor cartridges
-
We use syringes that can delivery as little as 0.2 microliters
of the sucrose/water USFnally, you'll need a good exhaust system to evacuate
the odor-laden air from the conditioning arena
-
If you would like more information about sources for these
items, or if you would like to obtain a copy of our software, contact Brian
H. Smith (smith.210@osu.edu)